Specifications

Carbon Dioxide Sensor Type Solid-state Infrared
Accuracy - Carbon Dioxide ± 1% of reading, or ± 0.2% CO2 (whichever is greater)
CO2 Minimum Detection Limit 0.1% (0-100% range) or 0.01% (0-20% range)
CO2 Calibration Controls Potentiometer SPAN adjustment for CO2 and zero adjusment
CO2 Calibration Frequency Every 12 months; calibration gas recommended.
Expected Lifetime (oxygen) 4-5 years
Oxygen Range 0.00 to 100%
Oxygen Sensitivity 0.1% O2
O2 Resolution 0.1% O2
Analog Output Not Available
Accuracy : +/- 1% of reading; +/- 0.1% absolute below 10.0%
Drift less than 0.4% per year
Oxygen Sensor Type Proprietary electrochemical cell, expected lifetime 4-5 years
Response Time 15 seconds to T95
O2 Sensor Lifetime 4-5 Years
O2 Sensor Body ABS Resin
Sample Connection Sample Probe
Sample Method Pump
Sample Pump Miniature diaphragm type with ~5cc/sec flow
Pump Timing 2-12 second adjustable pump time, potentiometer on side of analyzer
Minimum Sample Required 20cc
Sampling Assembly Probe with female luer connection for filter and needle, low volume PTFE tubing
Oxygen Calibration Controls SPAN adjustment, zero adjustment
Instrument Calibration Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on side (calibration gas recommended for CO2)
Exhaust port Vents to air
Power Supply Four "AA" NiMH rechargeable batteries; auto shut-off after 45 minutes idle. 100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request)
Backup Power Four "AA" alkaline batteries (not included)
Battery Operation Included
Input Voltage 100-240V, 50-60Hz
Battery Life 8-12 hours before recharging
Measures: O2, CO2 O2, CO2
Form Factor Portable
Size 8 x 4 x 1.8 in. (200 x 100 x 40 mm)
Weight 1 lb. (450 gms)
Warranty 2 years, parts and labor
Standards CE, RoHS
Included Consumables 200 foam septas, 2 luer fit needles, 2 particulate filters, 2 liquid block filters
Included Documentation Owners manual (includes operating instructions with detailed calibration instructions, calibration certficate, precautions), NIST calibration certificate
Origin of Goods Our products are manufactured in the U.S.A.
Products by Gas Oxygen and CO2
Products by Applications Gas Flow / Process Analyzers

What included

  • Built-in pump, sample probe.
  • Starter pack of 200 foam septas
  • 2 needles.
  • 2 particulate filters,
  • 2 moisture blocking filters,
  • Power supply / charger (for analyzers with optional internal battery) – (International compatible 100-240V)