Model 905V Pharmaceutical Vial O2 Analyzer |
||
|
|
|
|
LOW-VOLUME SENSOR DESIGN
The Model 905V sensor is designed specifically for this purpose. The oxygen sensor internal chamber has a very small volume of about 0.1 cc. The sensor is located directly behind the front panel to eliminate excessive volume. The sensor has in /out ports with a 0.03 in. ID. The vent port of the sensor connects to a short piece of narrow-bore tubing for sample exhaust and introduction of zero or calibration gases.
CALIBRATION SIMPLICITY
The oxygen sensor output is very linear through the entire range of measurement. This allows for a single point calibration to be performed anywhere in the measurement range. Span calibration is accurately done by testing room air, which should give a reading of 20.9% oxygen. Calibration standards can also be used if required by in-house test procedures.
LABORATORY OR PRODUCTION LINE
The Model 905 is a rugged benchtop unit housed in a heavy-duty metal enclosure, well suited for continuous use in routine QC testing in the laboratory. With built-in battery operation (optional) in addition to the AC mode, this unit can be easily transported and used at remote locations such as the packaging line.
DESIGN RELIABILITY
Microprocessor based electronics are used for sensor signal processing, battery charging and on/off control to provide high accuracy and reliability. The heavy duty, proprietary oxygen sensor is designed to provide many years of service, is totally sealed and requires no maintenance. The sensor output is compensated for ambient temperature variations.
All of our products are manufactured in the USA. |
||
![]() |